Ambulance (EMS) Medical Gas Outlets
Ambulance (EMS) Medical Gas Outlets

Ambulance (EMS) Hose Barb Medical Gas Outlets and Ambulance (EMS) DISS and NPT Medical Gas Outlets.

DOWNLOADS:
BROCHURESPEC SHEETS

Ambulance (EMS) Medical Gas Outlets Specs

FilenameSizeType
ape_sp_amb_ems_diss0.39MBPDF
ape_sp_hose_brb_ems_out0.37MBPDF

General Features

  • Compensates for installation depth variations
  • 360 degree swivel inlet connection
  • Gas specific indexing
  • 100% tested for flow, leaks and connector retention
  • Gas specific rough-in accepts either: Puritan-Bennett, Ohmeda, Chemetron, DISS, Oxequip/Medstar

Specifications

  • Hose Barb Outlets: The gas inlet tube of the outlet body shall be a hose barb fitting to permit connection of either 1/4" or 5/16" or 3/8" O.D. flexible hose with a crimp type ferrule.
  • DISS Outlets: Gas specific DISS connection brazed to the back body of the gas outlet which will allow a gas specific hose to be attached to the connection.
  • The Medical Gas Ambulance outlet shall be an Amico Alert-1 series.
  • Medical Gas Ambulance outlets shall be gas specific and permanently labelled for the services indicated.
  • All outlets meet applicable requirements of "Federal Specification for Ambulances" KKK-A-1822C. Amico ambulance outlets comply with NFPA-99.
  • For positive pressure gas services, the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall be rated at a maximum of 200 psi (1,379 kpa) in the event the latch valve assembly is removed for maintenance.
  • Outlet bodies shall be gas specific by indexing each gas service to a gas specific dual pin indexing arrangement on the respective identification module.
  • A large color coded front plate shall be used for ease of gas identification and aesthetic appeal.
  • The latch-valve assembly shall be Ohmeda Diamond Quick Disconnect, Chemetron Quick Connect, Puritan-Bennett Quick Disconnect, DISS compatible or Oxequip/Medstar compatible and accept only corresponding type gas specific adapters.
  • All outlets shall be cleaned and degreased for medical gas service, factory assembled and tested.
  • Amico products comply with NFPA-99.

Model Numbers:

 

O-FASC-QD-L-XXX



Latch-Valve Assembly:

QD = Ohmeda compatible

CH = Chemetron compatible

PB = Puritan-Bennett compatible

DI = DISS

OX = OXEQUIP/MEDSTAR compatible

NOTE:

For ceiling applications, only DISS outlets are recommended

 

The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

' Please see note above

 


 

O-BAKCOND-L-XXX (DISS connection)


 

The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

' Please see note above

 


 

O-BAKCONP-L-XXX (1/8” MNPT)


The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

' Please see note above

 


 

O-FASC-QD-L-XXX



Latch-Valve Assembly:

QD = Ohmeda compatible

CH = Chemetron compatible

PB = Puritan-Bennett compatible

DI = DISS

OX = OXEQUIP/MEDSTAR compatible

NOTE:

For ceiling applications, only DISS outlets are recommended

 

The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

***Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

*** = Available only in ¼”

' Please see note above

 


 

O-BAKCONB-L-XXX (1/4” Hose Barb connection)


 

The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

***Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

*** = Available only in ¼”

' Please see note above

 


 

O-BAKCONH-L-XXX (5/16” Hose Barb connection)


The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

***Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

*** = Available only in ¼”

' Please see note above

 

 


 

O-BAKCONG-L-XXX (3/8” Hose Barb connection)


The “L” defines the Language:

U = English (NFPA) Oxy, Air, Vac and WAG only

E = English (CSA/ISO)

F = French (CSA/ISO)

S = Spanish (NFPA) Oxy, Air, Vac and WAG only

 

The “XXX” defines the Gas:

Oxygen = OXY

Medical Air = AIR

Nitrous Oxide = N2O

*Nitrogen = NIT

** Carbon Dioxide = CO2

MedVac' = VAC

Waste Anesthetic Gas Disposal (NFPA) = WAG

Anesthetic Gas Scavenging System' (CSA) = AGS

***Instrument Air = INA

 

* = Available only in DISS

** = Not available for Puritan-Bennett

*** = Available only in ¼”

' Please see note above

 

Languages

  • English
  • Français