MEDICAL AIR TREATMENT SYSTEM WITH
DESICCANT AIR DRYERS
Furnish and install, where shown on the drawings a prefabricated Medical Air Treatment system with desiccant air dryers model 3DDF_____ as manufactured by:
Amico Source
Toll Free Tel: 1.877.462.6426
The unit furnished shall be
a standard catalog item of the supplier regularly engaged in the business of
providing packaged systems for hospitals and laboratories and shall meet and
exceed the requirements of NFPA 99.
The Amico Alert-2
Duplex Desiccant Air Dryer system shall be
a completely packaged NFPA 99 and NEC compliant assembly featuring redundant
twin tower desiccant air dryers with integral purge saver control, dual 0.5
micron pre-filters, dual 0.5 micron after-filters, line pressure regulating
valves, dew point monitor, CO monitor, test port, and other accessories required
to meet and exceed the current code requirements.
All components shall be pre-piped and pre-wired to single-point service connections in the field to eliminate field assembly errors, simplify installation and reduce installation costs. The only field connections are air intake, air discharge and power connection. All interconnecting copper/brass piping as well as wiring is completed and operationally tested prior to shipment.
Each unit shall consist of two identical banks of air treatment equipment, piped in parallel and provided with valves to by-pass either bank for filter element replacement, maintenance and repair work while still treating medical compressed air through the other bank without any sacrifice in air quality. Each bank consists of three stages of treatment.
The first stage is a
prime efficiency coalescer with particle removal down to 0.5 micron with
99.9999% retention. This filter removes aerosols and solid particles. The filter
is equipped with electronic drain and element change indicator.
The second
stage is a desiccant heatless air dryer. Each dryer is sized for 100% of the
system NFPA peak calculated demand. The dryer is designed to provide a maximum
dew point below the frost point of 0o C
(+32oF)
per NFPA 99, 2002 Edition. The dew point is attained by directing the flow of
saturated compressed air over desiccant which is contained in two separate but
identical "twin" towers. As the saturated compressed air flows though
the "on line" tower, its moisture adheres to the surface of the
desiccant. A solid state controller automatically cycles the flow of compressed
air between the towers. While the "on line" tower is drying, the
"off line" tower is regenerating. Built-in purge saver control will automatically minimize and
adjust the amount of purge air to match the variable air flow.
The third
stage is a prime efficiency particulate afterfilter with particle removal down
to 0.5 micron. The after filter element is provides high particle retention, low
pressure drop and long element life.
Downstream pressure
regulators will maintain constant discharge pressure of 55 PSIG (field adjustable).
Digital dew point and
CO monitors with alarm set points at +390F
and 10 PPM are provided with dry contacts for connection to remote alarm panels.
A "demand check" for maintenance is included per current NFPA 99 for
each instrument.
The system
includes a UL labeled control panel in a NEMA 12 enclosure. The panel includes a
power "On-off" switch and individual dryer dual control selector
switches for "Purge saver" and "Desiccant regeneration.
The Medical Air Treatment System and its component parts will undergo a complete electric and pneumatic test prior to shipment.
The Medical Air Treatment
System is guaranteed by the
manufacturer for a period of 12 months from the date of start-up or 18 months
from the date of shipment (whichever comes first) against defects in design,
materials, or construction.
The service of a factory
trained representative will be available at the jobsite to check installation,
start-up and instruct operating personnel in proper operation and maintenance.