LUBRICATED ROTARY VANE PUMPS
DUPLEX STACK MOUNTED VACUUM SYSTEMS
Furnish and install, where shown on the drawings a prefabricated duplex stack mounted lubricated rotary vane vacuum system model____________ as manufactured by:
Amico Source
Toll Free Tel: 1.877.462.6426
The unit furnished shall be a standard
catalog item of the supplier regularly engaged in the business of providing
packaged systems for hospitals and laboratories and shall meet and exceed the
requirements of NFPA 99.
The package shall include two lubricated
rotary vane vacuum pumps and associated equipment, one vertical ASME tank and
one duplex control panel. The only field connections required will be system
intake, exhaust and power connection at the control panel. All components shall
be completely pre-piped and pre-wired to single-point service connections. All
interconnecting piping and wiring shall be completed and operationally tested
prior to shipment. Provide liquid tight conduit, fittings and junction boxes for
all control and power wiring.
The medical vacuum pumps shall be of the
rotary vane air-cooled design with integral, fully recirculating oil supply with
sight gauge to indicate oil level. The oil separation system shall be integral
and shall consist of no less than four stages of internally installed oil and
smoke eliminators. This system shall be capable of removing 99.9+ percent of all
oil and smoke particles from the exhaust. Each pump shall include a built-in
anti-suck-back valve mounted at the pump inlet; and each pump shall be equipped
with three non-asbestos vanes, each having a minimum life of 30,000 to 40,000
hours.
Each vacuum pump shall be driven by a ____
HP, 3 phase, 60 cycle, ______ volt, 1750 RPM, TEFC NEMA C-face, foot mounted
motor. Each pump shall have a capacity of ____SCFM at 19 “HG.
The system shall include the following
accessories for each pump: inlet check valve, inlet isolation valve, vacuum
control switch, oil temperature gauge, thermal malfunction switch and vacuum
control switch. Provide flexible connectors on inlet and exhaust of each pump,
exhaust tee with union, drip-leg with cock valve as well as copper tubing with
shut-off cock for gauge and vacuum switches. The system shall include a _______
gallon vacuum storage tank of ASME construction. The tank shall be rated for
full vacuum service and shall be equipped with a valved by-pass, vacuum gauge
and manual tank drain. The inside of the tank shall be coated for rust
protection with a two component coating which provides a hard, durable lining.
Provide vibration mounting per NFPA 99.
The system shall include a UL listed control
panel in a NEMA 12 enclosure with the following accessories for each
pump:
Externally operable circuit breaker with
door interlock, control circuit transformer with fused primary and secondary
coils, H-O-A switch, magnetic starter with 3 leg overload protection, hour
meter, motor running light and
minimum run timer to prevent short cycle operation. Provide the panel with a
multiple position selector switch for selection of normal operation (automatic
alternation) or manual selection of lead and lag pumps if one of the pumps is
taken out of service due to scheduled maintenance. Provide audible and visual
local alarm (complete with indicating lights and individual sets of auxiliary
contacts wired to the terminal strip for remote alarm indication) for the
following: vacuum pump thermal malfunction and reserve vacuum pump in use.
Provide manual reset for thermal malfunction shut-down. All control and alarm
functions shall remain energized while any vacuum pump in the system remains
electrically on-line. The lag vacuum pump shall be able to start automatically
if the lead vacuum pump fails to operate.
The vacuum system shall be guaranteed in
writing by the manufacturer for a period of 12 months from the date of start-up
or 18 months from the date of shipment (whichever comes first) against defects
in design, materials, or construction. In addition, the bare pumps shall be
guaranteed for 36 months from the date of shipment.
The service of a factory trained
representative shall be made available at the jobsite to check installation,
start-up and instruct operating personnel in the proper operation and
maintenance.